• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATUN TIB SLV M/L 29MM FULL POR; KNEE TIBIAL TRAY ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATUN TIB SLV M/L 29MM FULL POR; KNEE TIBIAL TRAY ACCESSORY Back to Search Results
Catalog Number 151111201
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 05/22/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A manufacturing record evaluation (nc search) was performed for the finished device product code: 151111201, lot - m82p80 and no non-conformances / manufacturing irregularities were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device product code: 151111201, lot - m82p80 and no non-conformances / manufacturing irregularities were identified.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient was originally operated on for an infected non-union of fusion.Since cultures were negative, surgeon chose to reimplant the patient.Continued drainage suggests infection.Since the patient continues to drain despite negative cultures, the surgeon chose to remove all components and convert to a static spacer utilizing lps components, placement of resorbable antibiotic beads and high dose antibiotic cement to coat the components.All components were well fixed but lack of bony ingrowth also suggests persistent infection.An i&d was performed on (b)(6) 2025, where the poly and pin were replaced.Depuy 2 with gentamicin was likely utilized, as that is the surgeon's norm.No delay occurred and no patient harm was noted.Doi: (b)(6) 2025.Dor: (b)(6) 2025.Affected side: left knee.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATUN TIB SLV M/L 29MM FULL POR
Common Device Name
KNEE TIBIAL TRAY ACCESSORY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22214723
Report Number1818910-2025-09819
Device Sequence Number12801366
Product Code NJL
UDI-Device Identifier10603295436553
UDI-Public(01)10603295436553(17)20341231(10)M82P80
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number151111201
Device Lot NumberM82P80
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/22/2025
Supplement Date Received by Manufacturer06/18/2025
Initial Report FDA Received Date06/16/2025
Supplement Report FDA Received Date06/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
Patient Weight70 KG
-
-