• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG IPERIA 7 DR-T DF-1 PROMRI; ICD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG IPERIA 7 DR-T DF-1 PROMRI; ICD Back to Search Results
Model Number 392409
Medical Device Problem Code Premature Elective Replacement Indicator (1483)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
Device explanted due to eri indication with unexpected battery behavior.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Additional Manufacturer Narrative
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were reviewed.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.Next, the device was interrogated.The interrogation could be properly performed and revealed the eri battery status.Next, the icd was subjected to an analysis of the electrical parameters.The current consumption of the icd was checked and found to be normal and as expected.Analysis of the battery condition, however, showed an inconsistency of charge taken from the battery and the battery voltage.A premature battery depletion could be confirmed during analysis.Please note, this icd is affected by the field safety corrective action, bio-lqc, initiated in march 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPERIA 7 DR-T DF-1 PROMRI
Common Device Name
ICD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key22214859
Report Number1028232-2025-02892
Device Sequence Number8659361
Product Code MRM
UDI-Device Identifier04035479128777
UDI-Public(01)04035479128777(17)190731
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date07/31/2019
Device Model Number392409
Device Catalogue Number392409
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2025
Supplement Date Received by Manufacturer06/18/2025
09/08/2025
Initial Report FDA Received Date06/16/2025
Supplement Report FDA Received Date06/18/2025
09/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age65 YR
Patient SexMale
-
-