| Model Number |
1.11 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon visual inspection of the battery compartment, plastic case damage indicative of improper/forceful removal of the batteries with a foreign object was observed.Pts is confident that this damage occurred post-distribution, as there are no manufacturing steps that would cause this type of damage.The customer allegation of overheating was not observed on the returned analyzer.Numerous battery compartment improvements have already been implemented prior to these allegations from the field.Process improvements and part specifications have enhanced seating and hold capacity of the battery contacts.Additionally, a new battery compartment design (approved by the fda per k193406) eliminates the potential for overheating by removing the point of contact that an unseated terminal would reach.
|
| |
|
Event or Problem Description
|
|
Contact stated that when batteries are inserted into their cardiochek plus it gets hot to the touch.
|
| |
|