• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXFORD UNI TIB TRAY SZA RM; PROSTHESIS, KNEE, FEMOROTIBIAL, UNICOMPARTMENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXFORD UNI TIB TRAY SZA RM; PROSTHESIS, KNEE, FEMOROTIBIAL, UNICOMPARTMENTAL Back to Search Results
Catalog Number 154719
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Joint Dislocation (2374)
Date of Event 06/12/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10 ¿ 154912, oxf ph3 cementless fem sz xsm, lot 7237890.160792, oxf anat brg rt x-sm 5mm pma, lot 6786762.G2 ¿ foreign ¿ japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that the patient underwent revision surgery three months post implantation due to pain and bearing dislocated anteriorly.All implants were revised.No additional information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h2, h3, h6, h10, h11.No product was returned or pictures provided; a product evaluation could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.The reported event was confirmed by review of x-rays.Radiographs were provided and reviewed.The review identified postoperative short ap x-ray shows slight varus alignment of the tibial component and grossly neutral femorotibial knee joint alignment rather than standard normal valgus alignment of the knee.On later imaging, the bearing is anteriorly dislocated.It is uncertain if the mild relative varus alignment of the knee could have led to bearing wear predisposing to bearing dislocation.The device history record was reviewed and no discrepancies relevant to the reported event were found.Review of the complaint histories identified additional similar complaints for the reported items and no additional similar complaints for the reported part and lot combinations.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXFORD UNI TIB TRAY SZA RM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, UNICOMPARTMENTAL
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22218794
Report Number3002806535-2025-00238
Device Sequence Number8659456
Product Code NRA
UDI-Device Identifier05019279388851
UDI-Public(01)05019279388851(17)340621(10)J7792994
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number154719
Device Lot NumberJ7792994
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2025
Supplement Date Received by Manufacturer10/28/2025
Initial Report FDA Received Date06/17/2025
Supplement Report FDA Received Date10/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age78 YR
Patient SexFemale
Patient Weight44 KG
-
-