| Model Number |
24970A |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: performance data collected from the mobile programmer was received and analyzed.Analysis of the service logs found the customer comment that noise was observed with the mobile programmer was confirmed.It was also determined at analysis that the mobile programmer had a loss of connection.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the mobile programmer experienced a sensing issue with the right atrial (ra) lead, where noise was observed during passive analysis, making it difficult to assess the passive sensing on the lead.Troubleshooting steps were taken to include swapping out the lead for a new lead, but the second lead resulted in the same wide band of noise.It was further reported that the attempted and successfully implanted right atrial (ra) leads exhibited noise.One lead was attempted not used and replaced with a successfully implanted lead.Performance data from the mobile programmer application was received and it was determined at analysis that the mobile programmer application had a loss of connection.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|