It was reported that a regular full flap approach was performed to place the implant at the first molar (upper right) site, #16 with 8mm of bone under the sinus floor.The implant was placed right after a bone graft (internal sinus lift) was done; but it was a bit too long (into the sinus) and perforated the sinus, so it was removed and a 10mm implant was placed instead.
|
|
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.D9: device availability and return date were updated g3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 67.H10: narrative/data was updated.Zimvie received the reported implant for evaluation.Visual evaluation of the as returned device identified the implant with signs of use.However, no apparent damage / malfunction that would cause or contribute to the reported event was identified.Measurements match drawing.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are patient factors and improper techniques.Therefore, based on the available information, device malfunction did not occur.The reported event was non-verifiable with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|