• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMPL TAPERED SCR-V HA 6MM 5.7MM 10MM; TSV6H10

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMPL TAPERED SCR-V HA 6MM 5.7MM 10MM; TSV6H10 Back to Search Results
Catalog Number TSV6H10
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 08/20/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: weight unknown / not provided.G4: additional pma/510(k) number ¿ k011028|k013227.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported that the implant at tooth site #30 was removed due to bone loss.Clinically found a buccal fistula with probing depths and bleeding.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D9: device availability.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H10: additional narrative.Dhr review was completed for the subject lot number 63749192.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 63749192 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : medical : bone loss.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TAPERED SCR-V HA 6MM 5.7MM 10MM
Common Device Name
TSV6H10
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key22236589
Report Number0002023141-2025-01608
Device Sequence Number8976390
Product Code DZE
UDI-Device Identifier00889024019454
UDI-Public(01)00889024019454(17)220831(10)63749192
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K061410
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Expiration Date08/31/2022
Device Catalogue NumberTSV6H10
Device Lot Number63749192
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/17/2025
Supplement Report FDA Received Date08/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/15/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-