| Catalog Number |
778426 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, the patient underwent ureteroscopy for kidney stones.During the surgery, when the procedure reached the catheter placement stage, a ureteral stent was inserted.After unpacking the stent, a small notch was found at the end of the stent resembling a pig's tail for which actual sample and photos are available.After clinical assessment, a new package was chosen, and the product had no issues, allowing the subsequent procedures to be completed smoothly and the surgery to be successfully finished.There was a delay in the surgery, but no other impacts.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6)
2025, the patient underwent ureteroscopy for kidney stones.During the surgery, when the procedure reached the catheter placement stage, a ureteral stent was inserted.After unpacking the stent, a small notch was found at the end of the stent resembling a pig's tail for which actual sample and photos are available.After clinical assessment, a new package was chosen, and the product had no issues, allowing the subsequent procedures to be completed smoothly and the surgery to be successfully finished.There was a delay in the surgery, but no other impacts.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed cause unknown.Received (1) photo samples.The photo sample was returned and could not be evaluated for the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), used ureteral stent.Visual inspection of the sample noted that the ureteral stent was broken at the top by the pig tail.The labeling is found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Based on the results of the investigation, no additional actions can be taken at this time.Correction: d,e,h.Udi format updated with available product information per gudid.Initial reporter name: (b)(6).H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|