| Model Number |
630-2431211-01 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device failed self-test for defib function.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including defib cycling, discharging, and bench handling after both hot and cold soak, without duplicating the customer's report.An internal inspection found no discrepancies.The processor/bridge/pace board, ecg board and defib flex cable were replaced as a pre-caution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|