Model Number CDS0706-NT
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 05/23/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2025, a patient presented with grade 4 degenerative mitral regurgitation, a posterior calcium shelf with short restricted posterior leaflet and elevated starting gradient for a mitraclip procedure.During the procedure there was challenges in reducing the mr.A mitraclip xtw and a mitraclip nt were attempted but were not implanted.During the grasping attempts, there was a tissue perforation noted on the posterior leaflet.Per the physician, there were no device malfunctions.The final mr was grade 3-4.There were no clinically significant delays reported.An additional medical procedure has been scheduled.
 
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Based on the information reviewed, the reported unspecified tissue injury appears to be related to patient conditions (calcified, thin, short and restricted posterior leaflet).Tissue damage/injury is a known possible complication associated with mitraclip procedures.Unexpected medical intervention and hospitalization were a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Common Device Name
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key22237855
Report Number2135147-2025-03261
Device Sequence Number15728012
Product Code NKM
UDI-Device Identifier05415067037367
UDI-Public(01)05415067037367(17)251111(10)41112A1119
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/11/2025
Device Model NumberCDS0706-NT
Device Catalogue NumberCDS0706-NT
Device Lot Number41112A1119
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2025
Supplement Date Received by Manufacturer07/11/2025
Initial Report FDA Received Date06/17/2025
Supplement Report FDA Received Date07/16/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STEERABLE GUIDE CATHETER
Outcome Attributed to Adverse Event Other; Required Intervention; Hospitalization;
Patient SexUnknown
-
-