• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SAFESTEP; PORT ACCESS NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SAFESTEP; PORT ACCESS NEEDLE Back to Search Results
Model Number N/A
Medical Device Problem Codes Material Twisted/Bent (2981); Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the needle tip cannot be fully retracted after removing.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.The device has not been returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of difficulty advancing the safety mechanism was confirmed to be caused by a bent needle.However, the exact cause of the bent needle remains unknown.The product returned for evaluation was one 20g safestep infusion set w/o y-site.The returned product sample was evaluated and the needle was observed to be bent near the needle base.The following observations were noted during the sample evaluation: ¿ usage residue was seen on the sample which proved that the product had experienced at least some use ¿ the tip of the needle bevel appeared unremarkable an attempt to advance the safety over the needle tip was unsuccessful as the safety could not pass the bent region of the needle.Based on the evidence provided with the returned sample it is unknown when or how the bend occurred in the needle.Damage to the needle could have occurred at the manufacturing facility, during shipping, during use, or during storage of the product, and no evidence was observed which supported a specific root cause of the event.This complaint will be recorded for future trending and monitoring purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESTEP
Common Device Name
PORT ACCESS NEEDLE
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
maddy vincent
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key22247617
Report Number3006260740-2025-03970
Device Sequence Number9659234
Product Code FPA
UDI-Device Identifier00801741066221
UDI-Public(01)00801741066221(11)240315(17)261027(10)ASHXFC089
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K153440
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberLH-0032
Device Lot NumberASHXFC089
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2025
Supplement Date Received by Manufacturer08/07/2025
Initial Report FDA Received Date06/17/2025
Supplement Report FDA Received Date08/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/15/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-