(b)(6).Method: the subject device ojr410 optiflow junior 2 nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the information provided by the healthcare facility, previous investigations of similar complaints and our knowledge of the product.Results: the healthcare facility reported that the tubing of the subject device was detached from the cannula tube joint.There was no reported patient consequence.Conclusion: without the return of subject device for evaluation, we are unable to determine the exact cause of the reported fault.All optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.Samples are also taken and pull tested to check the tube tensile strength and the glue joint strength at the cannula/tube joint, as well as the swivel grip joint.The subject cannula would have met the required specification at time of production.The user instructions which accompany the subject device ojr410 optiflow junior 2 nasal cannula show in pictorial format the correct placement and fitting of the cannula, and provide the following warnings: "appropriate patient monitoring (e.G., oxygen saturation) must be used at all times.Failure to monitor the patient may result in loss of therapy, serious harm or death." "ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient." "do not stretch the cannula on application, this may cause increase pressure to the patient's skin.If necessary, the cannula may be repositioned.".
|