• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER AND PAYKEL HEALTHCARE; NASAL CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER AND PAYKEL HEALTHCARE; NASAL CANNULA Back to Search Results
Model Number OJR410
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Method: the subject device ojr410 optiflow junior 2 nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the information provided by the healthcare facility, previous investigations of similar complaints and our knowledge of the product.Results: the healthcare facility reported that the tubing of the subject device was detached from the cannula tube joint.There was no reported patient consequence.Conclusion: without the return of subject device for evaluation, we are unable to determine the exact cause of the reported fault.All optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.Samples are also taken and pull tested to check the tube tensile strength and the glue joint strength at the cannula/tube joint, as well as the swivel grip joint.The subject cannula would have met the required specification at time of production.The user instructions which accompany the subject device ojr410 optiflow junior 2 nasal cannula show in pictorial format the correct placement and fitting of the cannula, and provide the following warnings: "appropriate patient monitoring (e.G., oxygen saturation) must be used at all times.Failure to monitor the patient may result in loss of therapy, serious harm or death." "ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient." "do not stretch the cannula on application, this may cause increase pressure to the patient's skin.If necessary, the cannula may be repositioned.".
 
Event or Problem Description
A healthcare facility in united kingdom reported that the tubing of an ojr410 optiflow junior 2 nasal cannula was detached from the cannula tube joint.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER AND PAYKEL HEALTHCARE
Common Device Name
NASAL CANNULA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego vargas banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key22248281
Report Number9611451-2025-00561
Device Sequence Number13272553
Product Code BTT
UDI-Device Identifier09420012435743
UDI-Public(01)09420012435743(10)2103154057(11)240528(17)270528
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K222197
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 05/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOJR410
Device Catalogue NumberOJR410
Device Lot Number2103154057
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2025
Initial Report FDA Received Date06/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-