MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC Back to Search Results
Model Number CM-884
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2025
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a veterinary procedure, when pulling it out from the retainer and preparing for suturing, the thread broke with a popping sound.It was unrelated to the swage and the center.It was noted that multiple product had the same issue.Another suture was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
D10 concomitant product: cm-884 biosyn 0 vio 75cm gs22 x36 (lot#: d1m3006y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a veterinary procedure, when pulling it out from the retainer and preparing for suturing, the thread broke with a popping sound.It was unrelated to the swage and the center.It was noted that two sutures from the same box were opened and had the same issue.Another suture was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: b5, g3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d9 (return date), g1 (mfr contact first name, last name, postal code, email, phone number), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.Two devices and eleven sealed representative were available for evaluation.Visual inspection noted that one of the needles was returned attached to the suture and the other needle was disengaged from the suture.The sutures were returned broken into multiple segments.For representative samples, visual and functional evaluation found no notable condition related to reported condition.It was reported that the thread broke with a popping sound.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: progressive loss of tensile strength and eventual absorption of the absorbable sutures occurs by means of hydrolysis.This loss is triggered when exposed to the environment or clinical application.Damage during use may result in inadequate sample length to perform testing and/or deformation of the suture contributing to a loss in tensile strength.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22250345
Report Number9612501-2025-01548
Device Sequence Number12722939
Product Code GAM
UDI-Device Identifier10884521032125
UDI-Public10884521032125
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K000037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCM-884
Device Catalogue NumberCM-884
Device Lot NumberD1M3006Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2025
Supplement Date Received by Manufacturer06/02/2025
08/14/2025
Initial Report FDA Received Date06/17/2025
Supplement Report FDA Received Date06/24/2025
08/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-