| Model Number |
ADM06008013P |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a peripheral artery disease procedure to treat a calcified lesion in the left mid superficial femoral artery, the inpact admiral drug coated balloon 018 burst during the procedure.The common femoral artery was described as highly calcified, and vessel preparation was performed using a 7mm non-medtronic device.A non-medtronic inflation device was used with 50/50 contrast fluid for inflation.After the first drug coated balloon burst on initial inflation, a second inpact admiral drug coated balloon was used, which also burst.The devices were safely removed from the patient.The procedure was completed after the second balloon burst.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional infomration: the balloon burst below rbp.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a peripheral artery disease procedure to treat a calcified lesion in the left mid superficial femoral artery, the in-pact admiral drug coated balloon 018 burst during the procedure.The common femoral artery was described as highly calcified, and vessel preparation was performed using a 7mm non-medtronic device.A non-medtronic inflation device was used with 50/50 contrast fluid for inflation.After the first drug coated balloon burst on initial inflation, a second in-pact admiral drug coated balloon was used, which also burst.The devices were safely removed from the patient.The procedure was completed after the second balloon burst.No patient complications have been reported as a result of this event.Additional infomration: the second balloon burst also occurred on the first inflation at, below, or at an unknown rated burst pressure, leading to the procedure being aborted.A 6x45 non-medtronic sheath and 014x300 non-medtronic guidewire were used.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|