| Model Number |
31300 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Detachment of Device or Device Component (2907); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned on june 05, 2025, will be forwarded to the manufacturing site for investigation.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted unsolicited.This event is filed under internal complaint id (b)(4).
|
| |
|
Event or Problem Description
|
|
It has been reported during recent insulation testing, identified arcing in certain areas of the sheathing for the mini laparoscopic instruments.Interestingly, the arcing occurred in the same area across different sheaths within the same tray.
|
| |
|
Additional Manufacturer Narrative
|
|
The surface of the shaft shows some very deep scratches/damage, which may have been caused by use with trocars or other instruments.This event is filed under internal complaint id (b)(4).
|
| |
|
Search Alerts/Recalls
|