| Model Number |
J20007D |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Eye Injury (1845); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/22/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section a4: information was requested but it is unknown/not provided.Section d6a: if implanted, give date: not applicable, as the system is not an implantable device.Section d6b: if explanted, give date: not applicable, as the system is not an implantable device.Section h3: the field service engineer (fse) was onsite to evaluate the system.Flap thickness within specifications.No problem found.System performing to specifications.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
|
| |
|
Event or Problem Description
|
|
The complaint pertains to an intraoperative event of vertical gas breakthrough reported by the customer.While the first eye was successfully treated, the flap on the second eye could not be lifted due to the escape of gas bubbles.The certified surgeon, observed the gas breakthrough and sought clarification regarding the appropriate timeframe for performing the photorefractive keratectomy (prk) procedure on the affected eye.Our clinical application specialist (cas) explained that there is no specific recommendation for the timeframe, leaving the decision to the surgeon's discretion, and advised against lifting the flap in such situations.No intraoperative intervention necessary.Flap was not lifted.No medication outside the standard of care was prescribed.Patient outcome after the prk procedure was 20/30-2, with pin hole 20/20+1 on the first day following surgery.Note: this is report 1 of 2.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h11, corrected data: based on further review of the complaint file, it was noted that the device manufacture date was not accurate in the initial medwatch report.The following section has been updated accordingly: section h4: device manufacture date, dec 1, 2009 all pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
|
| |
|
Search Alerts/Recalls
|