Catalog Number 0279401200
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was smoke being emitted from the hand piece, it also overheated.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Additional Manufacturer Narrative
It was reported that the device was discarded and will not be returned.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: rf wand overheated probable root cause: heat can result from insufficient suction which can result from: * inadequate hospital suction, leak, bad connection to hospital suction line, * clogging caused by suction path blockage, lower electrode mass due to variation in electrode thickness or opening cut use errors nonbendable probe bent with probe bender, or a bendable probe bent with anything other than the stryker rf probe bender probe used as a tool for mechanical displacement of tissue or bone, or to pry or scrub probe handle is submerged in liquid or suction tube is cut pinch clamps left closed when suction is desired or open if suction is not being used incorrect fluid management the reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that there was smoke being emitted from the hand piece, it also overheated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERFAS ENERGY 90-S CRUISE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andy dobos
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key22284907
Report Number0002936485-2025-00553
Device Sequence Number21065344
Product Code GEI
UDI-Device Identifier07613327057034
UDI-Public07613327057034
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K160050
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0279401200
Device Lot Number25048AE2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2025
Supplement Date Received by Manufacturer08/11/2025
Initial Report FDA Received Date06/18/2025
Supplement Report FDA Received Date09/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-