| Catalog Number |
283419 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Event or Problem Description
|
|
Report 2 of 2 for (b)(4).It was reported that during a meniscal repair surgery, it was observed that debris fell off the aggressive 4.0mm 5pk device and was found on the blade.It was reported that the device was changed to another one to continue the surgery and the same problem happened again.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the complaint device was discarded by the customer, therefore not returned to j&j medtech orthopaedics and unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number (m2407005), and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: investigation summary according to the information received, it was reported that during the surgery, debris fell off was found on the blade.Changed another one to continue the surgery, the same problem happened again.No additional information could be provided.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection found that the device had the inner shaft and the outer shaft with friction signs.The inner shaft was nicked.No other anomalies could be observed.The overall complaint was confirmed as the observed condition of the aggressive 4.0mm 5pk would have contributed to the complained device issue.Based on the investigation findings, the potential cause can be traced to blade wear and degradation.As per instructions for use (ifu); excessive side loading may result in blade wear and degradation.Adequate suction is recommended to reduce wear and degradation of the device, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: a manufacturing record evaluation was performed for the finished device lot number (m2407005), and no non-conformance was identified.
|
| |
|
Search Alerts/Recalls
|
|