|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3505455BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
04/10/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 31-year-old caucasian female patient underwent a breast augmentation revision with a 455cc mentor memorygel breast implant and experienced a rupture on the right side post-operatively.As a result, the patient underwent explantation on (b)(6) 2025.
|
| |
|
Additional Manufacturer Narrative
|
|
On july 11, 2025, the mentor failure analysis lab has received the device for evaluation.On july 18, 2025, the evaluation for the device was completed.Device evaluation summary: visual analysis of the returned device revealed that no damage or anomalies were observed on the breast implant.Leak testing was performed, in accordance with mentor procedures, and no areas of gel exposure were detected during the analysis.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture status identified by mri evaluation includes both suspected ruptures, which are those ruptures identified by mri but not confirmed by explantation and examination of the device, and confirmed ruptures, which are those ruptures that are confirmed by evaluation of the explanted devices.The event described could not be confirmed as the breast implant was returned without detectable gel exposure.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: since no malfunction was observed during the investigation.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|