| Model Number |
694765 |
| Medical Device Problem Code |
Decreased Sensitivity (2534)
|
| Health Effect - Clinical Codes |
Dizziness (2194); Presyncope (4410); Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
05/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient presented to the emergency room due to dizziness/near syncope and congestive heart failure.It was noted that the right ventricular (rv) lead exhibited a decrease in r-wave amplitude noted on lead trends.The lead remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|