• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS, INC. ATELLICA CH CALCIUM_2 (CA_2)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS, INC. ATELLICA CH CALCIUM_2 (CA_2) Back to Search Results
Model Number N/A
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to siemens erroneously elevated calcium_2 (ca_2) results obtained on two (2) patient samples on an atellica ch 930 analyzer.The erroneously elevated patient results were reported to the physician and questioned.The same samples were reprocessed for one patient on the same atellica ch 930 analyzer and for the second patient on an alternate atellica ch 930 analyzer.Lower results were obtained, considered correct and reported to the physician.There are no known reports of patient intervention or adverse health consequences due to the erroneously elevated calcium_2 (ca_2) results.
 
Additional Manufacturer Narrative
An outside united states (ous) customer contacted the siemens healthcare diagnostics remote services center (rsc) regarding erroneously elevated calcium_2 (ca_2) results on two (2) patient samples obtained on an atellica ch 930 analyzer.Siemens is investigating the event.
 
Additional Manufacturer Narrative
Additional information (10-jul-2025): siemens healthcare diagnostics service operations support (sos) concluded the investigation of the event.Sos reviewed the information provided by the customer and the data from the instrument.No reagent or instrument issues were identified.Quality control (qc) recovered within the customer's laboratory ranges at the time of the event.Sos reviewed the calibration data for ca 2 on atellica ch 930 analyzer and observed variability for the c1 calibration coefficient across the different calibration events.The customer performed a new calibration prior to repeat testing which produced acceptable results.Based on the available information, the cause of the event maybe due to inconsistent reconstitution/handling of the calibrator material.A potential product issue was not identified.The customer is operational.The device is performing within specifications.No further evaluation is required.In section h6, the investigation finding and investigation conclusion codes were updated.Initial mdr 2432235-2025-00161 was filed on 18-jun-2025.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATELLICA CH CALCIUM_2 (CA_2)
Common Device Name
CALCIUM
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
511 benedict avenue
tarrytown NY 10591 5097
Manufacturer (Section G)
RANDOX LABORATORIES LTD
55 diamond road
registration # 8020890
crumlin, county antrim BT29 4QY
UK   BT29 4QY
Manufacturer Contact
eiman sulieman
500 gbc drive
po box 6101
newark, DE 19714-6101
7372808688
MDR Report Key22286686
Report Number2432235-2025-00161
Device Sequence Number8310125
Product Code CJY
UDI-Device Identifier00630414220697
UDI-Public(01)00630414220697(10)242373(17)251001
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K083386
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2025
Device Model NumberN/A
Device Catalogue Number11097644
Device Lot Number242373
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/05/2025
Supplement Date Received by Manufacturer07/10/2025
Initial Report FDA Received Date06/18/2025
Supplement Report FDA Received Date07/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/23/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-