|
The reported event could not be confirmed since the device was not returned for evaluation and found to be functional in nature.The returned set targeting arm was received in overall good condition, with slight marks.We also see that the marking color is moved from white to fawn, which is caused by a high number of sterilization and cleaning cycles.A functional inspection was conducted with a sample nail adapter, nail, sleeve, drills, and it was checked at the lateral, medial, and posterior sections, and it was observed that the locking was working as it was intended.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the above investigation, the root cause can be attributed user related, as in the functional inspection device was found to be functional.If any additional information is provided, the investigation will be reassessed.
|