| Model Number |
470309-14 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during central processing, the 8mm mega suturecut needle driver instrument had a broken cable.There were no reports of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive 8mm mega suturecut needle driver instrument to perform failure analysis.The instrument was analyzed and found to have damaged grip cable at distal end.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|