Concomitant products: 125028, 125028 28fr lt bronco cath pu cuff x1, (lot# 202008026x), 125028, 125028 28fr lt bronco cath pu cuff x1, (lot# 202006377x), 125028, 125028 28fr lt bronco cath pu cuff x1, (lot# 202007434x), 125028, 125028 28fr lt bronco cath pu cuff x1, (lot# 202304062x), 125028, 125028 28fr lt bronco cath pu cuff x1, (lot# 202008026x),125028, 125028 28fr lt bronco cath pu cuff x1, (lot# 202008026x), 125028, 125028 28fr lt bronco cath pu cuff x1, (lot# 202006377x), 125028, 125028 28fr lt bronco cath pu cuff x1, (lot# 202007434x), 125028, 125028 28fr lt bronco cath pu cuff x1, (lot# 202007434x), medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted a slit on each of the tracheal cuffs which measure 7.97mm, 10.21mm and 6.31mm in length respectively.Functional testing was done by an attempt made to inflate the returned units as per the parameters set out in if001 failed as the tracheal cuffs could not be inflated.It was reported that the cuff of the bronco catheter had an air leak.The reported issue was confirmed.The most likely cause was unintended use error caused or contributed to event.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: warnings / precaution (cuff-related): where it states ¿various bony anatomical structures (e.G., teeth) within the intubation route or any intubation tools with sharp surfaces present a threat to maintaining cuff integrity.Care must be taken to avoid damaging the thin-walled cuffs during intubation which would create the need to subject the patient to the trauma of extubation and re-intubation.If either cuff is damaged, the tube should not be used.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|