SMITH & NEPHEW, INC. JRNY II BCS CNSTRD ART ISRT 1-2 LT 9M; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
74029221 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Discomfort (2330); Joint Dislocation (2374)
|
| Date of Event |
05/29/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, after a tka surgery performed on (b)(6) 2024, a first revision surgery was performed on (b)(6) 2025 due to sinking of the tibial baseplate and discomfort ((b)(4)).A second revision surgery was performed on (b)(6) 2025 due to feeling of "wrongness" and posterior dislocation of tibia.The jrny ii bcs cnstrd art isrt 1-2 lt 9m was replaced for a journey ii deepdish insert.Surgeon mentioned that the patient had stiff joints and flexion contracture.When bending the knee, the hamstrings pushed the tibia backward and the femur forward, causing dislocation.The surgeon doesn't think products failure.The current health status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6 this complaint was opened by smith+nephew to document a patient complication reported that includes reference to the use of a smith+nephew product without evidence about a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Search Alerts/Recalls
|
|
|