MAUDE Adverse Event Report: TREACE MEDICAL CONCEPTS, INC. LAPIPLASTY 4.0MM FULLY THREADED SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TREACE MEDICAL CONCEPTS, INC. LAPIPLASTY 4.0MM FULLY THREADED SCREW Back to Search Results
Model Number SK38
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that after an initial bunion surgery on (b)(6) 2024, a broken screw was discovered via radiographic image from a post-op follow-up.No other patient outcomes were reported as a result of this event.No devices were returned for evaluation as all hardware currently remains implanted.The device history records were reviewed and no issues were identified during the manufacture and release of the devices that could have contributed to what was reported.A number of factors could have contributed to what was reported and although the most likely cause cannot be determined based on the available information and without the return of the device, the broken screw was likely subjected to excessive bending stresses which resulted in the breakage.The company will supplement this mdr as necessary and appropriate.
 
Event or Problem Description
It was reported that after an initial bunion surgery on (b)(6) 2024, a broken screw was discovered via radiographic image from a post-op follow-up.No other patient outcomes were reported as a result of this event.All hardware currently remains implanted.Although there has been no injury reported and this malfunction has not resulted in a revision surgery to date, there have been similar events where this malfunction has resulted in a revision surgery.Therefore, a mdr will be filed to ensure full compliance with 21 cfr part 803.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPIPLASTY 4.0MM FULLY THREADED SCREW
Common Device Name
SCREW
Manufacturer (Section D)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer (Section G)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer Contact
renae ginter
100 palmetto park place
ponte vedra, FL 
9043735940
MDR Report Key22289244
Report Number3011623994-2025-00064
Device Sequence Number8893478
Product Code HWC
UDI-Device Identifier00853114006303
UDI-Public(01)00853114006303(17)250621(10)300169704
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K242671
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 06/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2025
Device Model NumberSK38
Device Lot Number300169704
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2025
Initial Report FDA Received Date06/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexFemale
-
-