| Model Number |
DW OUTFLOW CASSETTE/TUBE SET |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
Swelling/ Edema (4577)
|
| Date of Event |
05/27/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a shoulder arthroscopy, an error message appeared with the dual console.Too much fluid entered the patient/shoulder too quickly.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully by changing to an open procedure.It was not necessary to do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6: the complaint allegation cannot be confirmed due to the inability to replicate the reported failure through functional testing.One unpackaged ar-6430 outflow tube set was received for investigation.The device was unable to be fully decontaminated and was evaluated via attachments.Visual evaluation of the attached photos noted no apparent issues with the ar-6430 outflow tube set.Functional testing was not performed due to the biocontamination hazard risk.The most likely cause for the reported failure can be attributed to use error due to incorrect connection of the main pump tubing during setup, likely the green and clear connectors were reversed on the pump.
|
| |
|