| Catalog Number |
284649 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Increase in Pressure (1491)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: investigation summary = > the product was not returned to j&j medtech orthopaedics, however a photo was provided for evaluation.Visual inspection of the returned photo found no observations pertaining to the nature of the reported event or any other anomaly.The overall complaint was not confirmed as the observed condition of the reuse patient set fms 24pk would have not contributed to the complained device issue.¿ based on the findings, the investigation could not draw any conclusions and no potential cause about the reported event can be established due to the insufficient evidence provided.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Event or Problem Description
|
|
It was reported that when using the reuse patient set fms, inflow tubing fms vue, and the fms vue ii pump-shaver box fluid management system devices, and system purging starts, an fc 04 error message started to appear, which refers that the pump had an excess of pressure, so the irrigation was paused.It was reported that the system was disconnected, and the tubes were re-installed, but the fc 04 error that the pump showed did not change.It was not reported if the event occurred during a surgical procedure.There were no reported adverse consequences to the patient.The exact date of the event was not reported, however, it was reported that the event occurred in 2025.No additional information could be provided.This is report 3 of 3 for (b)(4).
|
| |
|
Search Alerts/Recalls
|
|