| Model Number |
MA60AC |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that during an intraocular lens (iol) implant procedure he haptic broken during delivery of iol.The lens removed during initial procedure and replaced with another lens.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in d.9., h.3., h.6.And h.11.The empty lens case was returned in the carton.The lens was not returned for evaluation.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.The account indicated the use of a qualified cartridge and handpiece with a non-qualified viscoelastic.The root cause for the reported complaint may be related to a failure to follow the ifu.The viscoelastic indicated is not qualified for the lens or monarch combination used.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The instructions for use (ifu) instructs: use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.The trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.Haptic manufacturing is a validated process with multiple inspections to verify the quality of the produced haptic material.These inspections include a verification of the dimensional accuracy, tensile strength, flexibility and cosmetic appearance.The inspections ensure that the material meets all required specifications before it is allocated for use.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|