| Model Number |
PSG100 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to a pulsed field ablation procedure, an error message was received indicating that there was an embedded power-on-self-test (post) failure.The failure was verified and not resolved.Multiple troubleshooting steps were attempted, including rebooting the device several times and rebooting without the cables attached, but no resolution was found.The unit was onsite for evaluation.Technical services recommended replacement of the evaluation unit and depot repair of the broken unit.A backup device was used.Despite this, the case was aborted and the patient was under general anesthesia.No patient complications have been reported as a result of this event.
|
| |
|