Catalog Number 364953
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported prior to use of bd vacutainer® c&s transfer straw kit, an unspecified number of kits had the needle inside the straw separated from the straw, including: straws detaching exposing needles, extra straws with needles exposed in the package, and no straw at all in package but holder and tube.Additionally, there were reports of tubes not filling, bent/crimped straws, empty packages, broken plastic and powder inside package, and a double label.There was no health impact or consequences reported.
 
Additional Manufacturer Narrative
Investigation summary: bd received 11 photos for investigation, which show the separation of the straw/needle from the holder but do not reveal the customer's indicated failures of no fill/no vacuum, damage, incorrect count, incorrect/missing label info, and fm.Additionally, 10 retained samples were visually inspected, and all components, including the stopper and straw/needle assembly, were found to be correctly assembled with no damage or fm observed, and the labels were correctly placed.Only 10 retained samples are kept for this product; therefore, the incorrect count could not be further investigated.Furthermore, a functional test for draw volume was conducted on 5 retained samples, and all tubes met the specification limits with no issues identified.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: separates.This complaint has not been confirmed for the indicated failure modes: damaged, draw volume(no fill), foreign matter, incorrect count, and label issues.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Event or Problem Description
It was reported prior to use of bd vacutainer® c&s transfer straw kit, an unspecified number of kits had the needle inside the straw separated from the straw, including: straws detaching exposing needles, extra straws with needles exposed in the package, and no straw at all in package but holder and tube.Additionally, there were reports of tubes not filling, bent/crimped straws, empty packages, broken plastic and powder inside package, and a double label.There was no health impact or consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® C&S TRANSFER STRAW KIT
Common Device Name
CULTURE MEDIA, NON-PROPAGATING TRANSPORT
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key22292973
Report Number1917413-2025-00806
Device Sequence Number12878414
Product Code JSM
UDI-Device Identifier30382903649533
UDI-Public(01)30382903649533(11)240403(17)251031(10)4094211
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
K024240
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/31/2025
Device Catalogue Number364953
Device Lot Number4094211
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2025
Supplement Date Received by Manufacturer07/30/2025
Initial Report FDA Received Date06/19/2025
Supplement Report FDA Received Date08/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/03/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-