• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD VERSAONE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD VERSAONE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number RNONB8STF
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10 concomitant products: mrasc0005 tower 240v mrasc0005 - (serial# (b)(6); rnonb8stf postion no blade 8mm std fix - (lot#unknown); rnonb8stf postion no blade 8mm std fix - (lot#unknown); rnonb8stf postion no blade 8mm std fix - (lot#unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that, during a robotic laparoscopic prostatectomy and lymphadenectomy procedure, while placing the trocars in the patient while they were already under anesthesia but prior to the start of the operation, it was found that the entire trocar shaft was within the abdominal wall, the tip was barely protruding from the cavity, and the upper part of the trocar had low mobility and was in contact with the skin.When the patient was placed in the trendelenburg position, the patient could not support the position for long.There weren't any long trocars, so the surgery was cancelled because they could not afford complications with this patient.The surgery will be performed laparoscopically at a later date without robotic support.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAONE
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key22293450
Report Number9612501-2025-01567
Device Sequence Number20949092
Product Code GCJ
UDI-Device Identifier10884521717640
UDI-Public10884521717640
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K162584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRNONB8STF
Device Catalogue NumberRNONB8STF
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2025
Initial Report FDA Received Date06/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-