| Catalog Number |
788426 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the surgery, when the ureteral stent was opened, it was found to be cracked, and then a replacement of the same specification product was used to complete the surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during the surgery, when the ureteral stent was opened, it was found to be cracked, and then a replacement of the same specification product was used to complete the surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Complete initial reporter's facility name: (b)(6).The reported event was confirmed cause unknown.Received (2) photo samples.The photo sample was returned and could not be evaluated for the reported failure.Visual evaluation of the returned sample noted one opened (with original packaging), ureteral stent kit.Visual inspection of the sample noted a break in the pig tail.The labeling is found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Based on the results of the investigation, no additional actions can be taken at this time.Corrections: d, e, h.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|