| Model Number |
436-0360 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
05/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The implant has not returned for evaluation.Photographs were provided which confirm the event of an inferior screw backing out of the plate.A review of the device history records could not be performed as the identifying lot number was not provided.Alphatec spine is submitting this report to comply with the fda regulations 21 cfr 803.Alphatec spine has made reasonable efforts to provide as much relevant information as available.Any field that is left blank was not known at the time of this submission.If additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
A patient underwent an anterior cervical discectomy and fusion spinal procedure on (b)(6) 2025.About 3 months postoperatively, a radiograph image revealed the screw was starting to backout of the plate.
|
| |
|