• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number MZ9265
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd maxzero multi-fuse extension set with needleless connector had leakage.The following information was provided by the initial reporter: the connection started leaking as soon as the ivf and nahco3 infusion was started.
 
Additional Manufacturer Narrative
One sample was received for quality evaluation.Visual inspection indicates that the luer adapter cracked.With the luer adapter being cracked.Leakage was identified coming from the crack in the luer.The customer complaint of leakage was confirmed.The investigation was forwarded to the manufacturing team and it was determined that the probable root cause was due to an over tightening or use of alcohol or antiseptic on the joining surfaces, and not allowing for the alcohol or antiseptic to dry before connection.This can cause cracks in the material.A device history record review for material# mz9265 and lot# 24119278 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 15nov2024.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key22293828
Report Number9616066-2025-01619
Device Sequence Number15672421
Product Code FPA
UDI-Device Identifier10885403240775
UDI-Public(01)10885403240775(11)241115(17)271115(10)24119278
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K140831
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMZ9265
Device Lot Number24119278
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/28/2025
Supplement Date Received by Manufacturer08/25/2025
Initial Report FDA Received Date06/19/2025
Supplement Report FDA Received Date08/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/15/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-