• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG 1788 CAMERA CONTROL UNIT CCU; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG 1788 CAMERA CONTROL UNIT CCU; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 1788010000
Medical Device Problem Codes No Display/Image (1183); Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported by the customer that the 1788 ccu unit was not working smoothly, with the camera going black unpredictably.The stryker rep further reported that he tried to troubleshoot the unit, but the issue still occurred.
 
Additional Manufacturer Narrative
Three attempts have been made to retrieve the device for evaluation, and it is yet to be received in-house.No further attempts to retrieve the device are required.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: camera going black unpredictably probable root cause: faulty solder joints in video path.Faulty prism board or connectors.Faulty xboard or ch cable connectors.Break in ch cable core.Faulty hdmi cable.Faulty ccu power supply.Internal ccu cable failure - power and/or communication.Improper/no fuse installed.Ac inlet board.Main board - video processor.Digital board - fpga.Transition board - coax connector.Electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge.Poor system grounding.Improper ccu timing.The reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported by the customer that the 1788 ccu unit was not working smoothly, with the camera going black unpredictably.The stryker rep further reported that he tried to troubleshoot the unit, but the issue still occurred.
 
Additional Manufacturer Narrative
D4 - serial number was updated.
 
Event or Problem Description
It was reported by the customer that the 1788 ccu unit was not working smoothly, with the camera going black unpredictably.The stryker rep further reported that he tried to troubleshoot the unit, but the issue still occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG 1788 CAMERA CONTROL UNIT CCU
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andy dobos
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key22294263
Report Number0002936485-2025-00563
Device Sequence Number15570195
Product Code GCJ
UDI-Device Identifier07613327580013
UDI-Public07613327580013
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K230605
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1788010000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/23/2025
Supplement Date Received by Manufacturer07/29/2025
10/06/2025
Initial Report FDA Received Date06/19/2025
Supplement Report FDA Received Date08/22/2025
10/31/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/01/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-