|
A field service engineer (fse) was at the customer site to address the reported event.The fse removed the sampling needle and found the tip was bent.Fse also found the line filter tube twisted and nearly kinked, likely caused by the customer replacing the filter element without disconnecting the tubing first.The needle wash block was intact.The fse could not determine the cause of the bent sample needle tip.The fse replaced the sampling needle, verified alignment, replaced the line filter tubing, cleaned and lubricated the sampling z-axis guide rod and lead screw due to grease buildup.The pump pressure was stable after tubing replacement, the alignment was confirmed during precision runs, calibrations, and patient precision tests were within acceptable ranges.No further action required by field service.The g8 analyzer is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(6) from 30apr2024 through aware date 30may2025.There were two similar complaints identified for bent sample needle during the search period including this case and no similar complaints identified for discrepant result.The g8 variant analysis mode operator¿s manual v 3.4 under chapter 1: introduction and applications: limitations of the procedure: total area dilution studies demonstrate that the assay is linear from a total area of 500 to 4000.However, the optimum total area is 700 to 3000.Abnormal red cell survival: the life span of red blood cells is shortened in patients with hemolytic anemias, and the actual life span depends upon the severity of the anemia.As a consequence, specimens from such patients may exhibit decreased glycohemoglobin levels compared to patients with normal red cell life span.The life span of red blood cells is lengthened in polycythemia or post-splenectomy patients.Specimens from such patients may exhibit increased glycohemoglobin levels.Hemoglobinopathies: the most commonly observed hemoglobin variants are hbs, hbc, hbd and hbe.These variants in their heterozygous state elute after the hba0 peak in their designated windows: hbs (h-v1 peak), hbd (h-v0 peak) and hbc (h-v2 peak).The hbe (p-hv3 peak) appears between sa1c and hba0 peaks.In all these cases, the sa1c% is reportable when these hemoglobin variants are present in the heterozygous state.When either of these hemoglobin variants is identified, the sa1c% is calculated in a proprietary way, depending on the type of hemoglobin variant, based on the area of sa1c, the area of identified hemoglobin variant and other peaks.When a p-hv3 peak is detected, a flag 43 is also reported.Glycemic monitoring for any patients displaying any homozygous hemoglobin (other than hbaa) such as hbss, hbcc or the double heterozygous sc, cannot be performed using sa1c because there is no hemoglobin a present.Alternative testing is mandatory for these types of patients.The most probable cause of the reported discrepant results was not established.Cause of the bent sample needle tip was not established; twisted line filter tubing.
|
|
A customer reported ¿discrepant patient result for hba1c¿ on the g8 analyzer.The customer stated the initial patient sample was 6.8%, which was reported out and the physician questioned the result; the patient is a known non-diabetic.The same sample tube was rerun and resulted with 5.4% percent, within the expected range for this patient (4.0%-6.0%).No impact to patient care due to the discrepant result.The customer also observed that the sample needle appeared to be bent.The customer is requesting service to inspect the g8 analyzer.A field service engineer (fse) was dispatched for further investigation.There is no indication of any patient intervention or adverse health consequences due to the reported discrepant patient result.
|