• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION BIPAP ST30; VENTILATOR, CONTINUOUS NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION BIPAP ST30; VENTILATOR, CONTINUOUS NON-LIFE-SUPPORTING Back to Search Results
Model Number DSX1030S11
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2023
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice/recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information that the patient alleging particles in device and airway.There was no report of medical intervention.The manufacturer was made aware of this complaint through a representative of the customer.The device was returned to the third-party service center for further evaluation.During the evaluation of the device at the third-party service center, the device was visually inspected and found no evidence of foam degradation.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the third-party service center.There was zero error code found.The third-party service center concludes that there was no visible foam degradation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION BIPAP ST30
Common Device Name
VENTILATOR, CONTINUOUS NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key22294637
Report Number2518422-2025-047590
Device Sequence Number8271286
Product Code MNS
UDI-Device Identifier00606959424500
UDI-Public(01)00606959424500
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K102465
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 06/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberDSX1030S11
Device Catalogue NumberDSX1030S11
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2023
Initial Report FDA Received Date06/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/24/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1973-2021
Patient Sequence Number1
Patient SexUnknown
-
-