Model Number SCORPION-MULTIFIRE NEEDLE
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 28th may 2025, it was reported by an arthrex employee via email that for an ar-13995n multifire¿ scorpion¿ needle, the scorpion needle tip broke.This was detected during use in an open rotator cuff repair procedure on (b)(6) 2025.The case was completed successfully as a new needle was opened.29 may 2025 - the complaint initiator updated that the needle was being fired by ar-13999mf.Whilst the arthrex employee was in theatre the scorpion needle tip was not retrieved (could not be found) however the surgeon confirmed that it was not inside the patient.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used during firing of the needle.Dfu-0392-sub-eo warns against the usages listed to help avoid needle breakage and potential patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCORPION-MULTIFIRE NEEDLE
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22296921
Report Number1220246-2025-02582
Device Sequence Number18259610
Product Code MDM
UDI-Device Identifier00888867015951
UDI-Public(01)00888867015951(11)250220(17)300131(10)15392935
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSCORPION-MULTIFIRE NEEDLE
Device Catalogue NumberAR-13995N
Device Lot Number15392935
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2025
Supplement Date Received by Manufacturer10/08/2025
Initial Report FDA Received Date06/20/2025
Supplement Report FDA Received Date10/08/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-