• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL+ UD 18IN 5-0 S/A P-3 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL+ UD 18IN 5-0 S/A P-3 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCP493H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 6/20/2025.H6 component code: g07002 - no device problem found.H6 component code: c22 - photo analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: if possible, could you kindly provide the lot number for the 1696g involved, please? the lot number for the 1696g involved is unknown h3 investigational summary: the product was returned to ethicon for evaluation.One empty labeled winding former, a needle and one used suture were received for analysis.Product code vcp433.The swage and attachment area were noted as expected during the visual inspection of the needle.The barrel hole of the needle was examined under magnification and remnant suture was noted.The suture was examined, and the end was observed to be broken.This type of detachment mode is most likely related to improper tipping, as the suture is breaking away from the swage area.A photo was provided and a foil and two sutures were observed inside the plastic bag.Based on the information currently available, damaged suture was identified during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2025 and suture was used.It was reported that the problem of pulling off suture needle happened in surgery.Changed another one to continue the surgery, the same problem happened.Changed another one to complete the surgery.There were no patient consequences reported.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 18IN 5-0 S/A P-3 PRM MP
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key22297130
Report Number2210968-2025-07137
Device Sequence Number8628989
Product Code GAM
UDI-Device Identifier10705031150034
UDI-Public(01)10705031150034(17)290930(10)103Z90
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCP493H
Device Lot Number103Z90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2025
Initial Report FDA Received Date06/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-