Product complaint # (b)(4).Date sent to the fda: 6/20/2025.H6 component code: g07002 - no device problem found.H6 component code: c22 - photo analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: if possible, could you kindly provide the lot number for the 1696g involved, please? the lot number for the 1696g involved is unknown h3 investigational summary: the product was returned to ethicon for evaluation.One empty labeled winding former, a needle and one used suture were received for analysis.Product code vcp433.The swage and attachment area were noted as expected during the visual inspection of the needle.The barrel hole of the needle was examined under magnification and remnant suture was noted.The suture was examined, and the end was observed to be broken.This type of detachment mode is most likely related to improper tipping, as the suture is breaking away from the swage area.A photo was provided and a foil and two sutures were observed inside the plastic bag.Based on the information currently available, damaged suture was identified during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.
|