| Model Number |
COMPR FT SCRW,2.5 MICRO,36MM LGTH |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 06/02/2025, a sales representative reported via email an issue involving an ar-8725-36h micro compression ft during a dip fusion procedure on (b)(6) 2025.While inserting the 2.5 headless compression screw, the tip of the cannulated driver broke within the head of the screw, and the distal tip of the screw fractured during insertion.Additional information requested on 6/18/2025.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
Additional information was received on 06/24/2025: fragments from the tip of the ar-8725-36h micro compression ft screw were retrieved from the patient using suction.The driver used with the screw for insertion during the event was ar-8737-37 driver shaft.The procedure was completed by implanting a shorter screw, the ar-8725-34h micro compression ft, and utilizing an ar-8737-37 driver shaft.The case experienced a delay of approximately 30 minutes to one hour.No additional anesthesia was administered to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.- one unpackaged ar-8725-36h, micro compression ft¿, titanium, 2.5 mm x 36 mm, batch 129038, was received for investigation.- upon visual inspection, it was noted that the distal and proximal ends of the screw were damaged/stripped and the tip of the screw was broken.No fragments were returned for investigation.- functional testing was not performed due to the damage to the device.- the most likely cause of the reported failure is user error, which can be attributed to over-torquing/over-engaging the driver within the screw.-refer to the investigation photos.-complaint allegation is confirmed.
|
| |
|