• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SALUDA MEDICAL PTY LTD EVOKE CAP12 PERCUTANEOUS LEAD KIT - 90CM (US); SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SALUDA MEDICAL PTY LTD EVOKE CAP12 PERCUTANEOUS LEAD KIT - 90CM (US); SCS LEAD Back to Search Results
Model Number 102879
Medical Device Problem Codes Inappropriate/Inadequate Shock/Stimulation (1574); Migration (4003)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 05/30/2025
Type of Reportable Event Serious Injury
Event or Problem Description
A patient implanted with an evoke spinal cord stimulation (scs) system reported a change in stimulation.The patient had reported adequate pain relief previously.An x-ray and impedance check were performed with no device issues identified.The patient reported previous falls; the evoke scs didn't cause or contribute to the patient¿s falls.The physician assessed that the patient¿s falls may have damaged the evoke closed loop stimulator (cls) and contributed to the changes in stimulation.A revision procedure (reported under manufacturer report number: 3021836309-2025-00161) was performed to remove the evoke cls and connect the leads to an evoke external closed loop stimulator (ecls).The stimulation change was not resolved following the revision procedure.Further troubleshooting performed, with patient leads connected to the evoke external closed loop stimulator (ecls), identified lead migration contributing to the reported event.An explant procedure was performed to resolve the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOKE CAP12 PERCUTANEOUS LEAD KIT - 90CM (US)
Common Device Name
SCS LEAD
Manufacturer (Section D)
SALUDA MEDICAL PTY LTD
5 eden park drive
macquarie park, nsw 2113
AU  2113
MDR Report Key22298585
Report Number3016111136-2025-00089
Device Sequence Number12902329
Product Code LGW
UDI-Device Identifier09352307001657
UDI-Public09352307001657
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/20/2025
Report Date (Section F) 05/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number102879
Device Catalogue Number3009
Device Lot Number9017195906
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date05/30/2025
Event Location Hospital
Date Report to Manufacturer05/30/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date06/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-