• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL; INTRAOCULAR LENS Back to Search Results
Model Number SY60WF
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lens was returned for evaluation.Solution is dried on the lens.The lens is cut in half, typical of insertion and removal from the eye.One haptic is broken in the gusset area and returned in the lens case.One half of the optic is broken into two pieces.Numerous scratches are observed on the optic portions of the lens.Cracks are also observed on the optic.Product history records were reviewed and the documentation indicated the product met release criteria.Qualified associated products were indicated.Lens damage was observed.All lenses are 100 percent inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company¿s release criteria.Based on our observation, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.Qualified associated products were indicated.The instructions for use instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ophthalmic viscosurgical device (ovd)-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.Ifu: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the intra ocular lens (iol) to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported that, during intraocular lens implantation surgery, the lens was cracked.The lens was inserted and removed during the same procedure.Additional information was received statin that the lens was cracked sometime between loading and insertion.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22299150
Report Number1119421-2025-01719
Device Sequence Number18327504
Product Code HQL
UDI-Device Identifier00380652251495
UDI-Public(01)00380652251495(17)270603(11)230605(10)15644742
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSY60WF
Device Catalogue NumberSY60WF.205
Device Lot Number15644742
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/28/2025
Initial Report FDA Received Date06/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/05/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON MONARCH IV INJECTOR; MONARCH III CARTRIDGE D; PROVISC VISCOSURGICAL DEVICE
Patient SexUnknown
-
-