• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3542501
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Capsular Contracture (1761); Cyst(s) (1800); Deformity/ Disfigurement (2360)
Date of Event 02/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a 51-year-old hispanic female patient underwent a breast augmentation revision with two unspecified gel breast prostheses and experienced bilateral ruptures post-operatively, which was diagnosed mammogram and ultrasound.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.This report is for the left side device.
 
Additional Manufacturer Narrative
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Since no lot number was provided, no manufacturing record evaluation review could be performed.Reason for device explant and/or reoperation: rupture.D4: udi: as the device information was not provided, the udi is not available.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On august 04, 2025, mentor received additional information regarding the patient's impacted device.It was determined that the impacted device was a 250cc mentor memorygel breast implant (catalog number 3542501) with lot number 6711514 and serial number (b)(6).It was also reported that the patient was to undergo device explantation on (b)(6) 2025.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On august 22, 2025, mentor received additional information regarding the patient's replacement device.It was reported that the patient's replacement device was a 275cc mentor memorygel breast implant (catalog number 3502751bc) with lot number 2098358.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On september 16, 2025, mentor received additional reporting that the patient also experienced capsular contracture (baker grade 4).The patient had firmness on the breast.On september 22, 2025, mentor also received additional information reporting that the patient had a simple cyst measuring 3 millimeters.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On october 05, 2025, a photo evaluation for the device was completed.Photo evaluation summary: upon visual evaluation of the image provided in the complaint, the implant was observed ruptured.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet a breast cyst is a fluid-filled sac within the breast.One breast can have one or more breast cysts.They are often described as round or oval lumps with distinct edges.Breast cysts are usually and typically do not require treatment, but they can be drained using fine needle aspiration if the cyst is large or uncomfortable.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.On october 14, 2025, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On november 13, 2025, the evaluation for the device was completed.Device evaluation summary: visual analysis of the returned device revealed that the breast implant was found to have an area of siltex cracking on the posterior view.In addition a tear was noted within the siltex cracking measuring approximately 3.1 cm.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.A breast cyst is a fluid-filled sac within the breast.One breast can have one or more breast cysts.They are often described as round or oval lumps with distinct edges.Breast cysts are usually and typically do not require treatment, but they can be drained using fine needle aspiration if the cyst is large or uncomfortable.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key22302392
Report Number1645337-2025-06877
Device Sequence Number8594192
Product Code FTR
UDI-Device Identifier00081317000846
UDI-Public(01)00081317000846(17)180430(10)6711514
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2018
Device Catalogue Number3542501
Device Lot Number6711514
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/30/2025
Supplement Date Received by Manufacturer08/04/2025
08/22/2025
09/16/2025
10/05/2025
11/13/2025
Initial Report FDA Received Date06/20/2025
Supplement Report FDA Received Date08/21/2025
09/09/2025
09/30/2025
10/27/2025
11/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age51 YR
Patient SexFemale
Patient EthnicityHispanic
-
-