| Model Number |
001366 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Sunmed holdings llc.Received a single report that referenced three different incidences, which were associated with separate units, involving three different patients.This is the third of three reports.Refer to 3004748541-2025-00011 for the first report, refer to 3004748541-2025-00012 for the second report.It was reported, the tube kinked which resulted in a non-delivery of oxygen.There was no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The product involved in the report has not been returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 20 jun 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d4; h4.Additional information-investigation conclusion: d4; h6; h11.The actual complaint product was not returned for evaluation; however, the reporter provided photographic/videographic evidence of the reported malfunction; evaluation of this evidence confirmed the complaint as reported.Product verification testing was performed and confirmed reliable oxygen delivery under kinked conditions.The device history record for lot 536539 was reviewed and the product was produced according to product specifications.All information reasonably known as of 20 aug 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
Sunmed holdings llc.Received a single report that referenced three different incidences, which were associated with separate units, involving three different patients.This is the third of three reports.Refer to 3004748541-2025-00011 for the first report.Refer to 3004748541-2025-00012 for the second report.It was reported, the tube kinked which resulted in a non-delivery of oxygen.There was no reported injury.
|
| |
|
Search Alerts/Recalls
|