Model Number DSX900T11
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
A device was returned to a third-party service center/manufacturer in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the third-party service center, the device was visually inspected and found evidence of foam degradation.During the evaluation, foam particles were observed.The device was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION
Common Device Name
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 
4125423300
MDR Report Key22302987
Report Number2518422-2025-106772
Device Sequence Number20837355
Product Code MNS
UDI-Device Identifier00606959424371
UDI-Public(01)00606959424371
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K090539
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 06/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberDSX900T11
Device Catalogue NumberDSX900T11
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2025
Initial Report FDA Received Date06/20/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/10/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-