This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 10, 4109, 4111 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4310 and 4315.H10: narrative/data was updated.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.As documented in the summary investigations, contributing factors for the reported event likely exist outside of zimvie control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported events remain non-verifiable as they are a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for infection and/or bone loss, problems reported in association with implant failure have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.
|
|
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H6: type of investigation codes were updated: 3331, 4109, and 4111.H10: narrative/data was updated.Zimvie received the reported device for evaluation.Visual evaluation was performed, the implant had signs of use with debris on its external threads.There was an implant removal tool stuck inside the implant.The irt device did not contribute to the sinus perforation.The irt device being stuck inside the implant occurred during removal however, there was no damage identified or signs of malfunction that would have contributed to the event.Measurements match drawing.Irt device recorded as a concomitant medical device.Device history review (dhr) review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.The customer did not submit images for the reported event.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was inadequate treatment planning and patient factors.Refer to attached summary investigation for infection and bone loss.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable with all the available information provided.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.Similar complaints for infection and / or bone loss related problems have been previously investigated.Refer to attached summary investigation.Investigations performed for hundreds of events where either of these conditions, infection, or bone loss, or both, have been reported, have concluded that the likely cause(s) for the implant failure in relation to these conditions are external factors.Those include medical conditions (e.G., diabetes, bruxism, etc.), patient habits (e.G., smoking, bad oral hygiene routine) and user error (e.G., surgical technique).There has not been any instance where either infection and/or bone loss, have resulted from a manufacturing or design defect.Furthermore, the probability of manufacturing or design defects that might lead to infection and result in bone loss occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming product.The analysis and result of investigations and the analysis of probability previously described are contained in the summary investigation reports performed for bone loss and infection, which are attached.Additionally, all device history record reviews verified that each implant was sterilized per procedure for every device.All complaint data used for the summary investigation was found to be conforming and did not meet capa/hhe/d/ or any further escalations.Therefore, there were no complaints which confirmed a manufacturing or design related issue that did or could cause or contribute to the reported event.
|