• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DJO SURGICAL; EMPOWR ACETABULAR SYSTEM, CUP, HEMISPHERICAL, MULTI HOLE, 64

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DJO SURGICAL; EMPOWR ACETABULAR SYSTEM, CUP, HEMISPHERICAL, MULTI HOLE, 64 Back to Search Results
Catalog Number 940-03-64J
Medical Device Problem Codes Unstable (1667); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 05/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The agent reported "(cup spun out.)".The previous surgery and the surgery detailed in this event occurred 3 months and 1 week apart.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.First, there was an ncmr#58887 associated with the main part # 940-03-64j, empowr acetabular system, cup, hemispherical, multi hole, 64, which document that out of 3 parts lot, all items were rejected due to the parts were manufactured to previous revision with.0004" roundness.New revision has increased roundness tolerance.Later, the rejected parts were reworked and retouched after proper justification through uai.Second, there was an ncmr#74301 associated with the concomitant part # 940-00-020, empowracetabular,bone screw,20mm which document that out of 50 parts lot, all items were rejected due to missing empower acetabular, screw product label 0400-0280 on dhr label record form.Quantity picked signed off on picklist.Later, the rejected parts were reworked and retouched after proper justification.All other items in the lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.The root cause of this complaint was a revision surgery due to dissociation.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities or trauma.
 
Event or Problem Description
Revision surgery - due to dissociation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
EMPOWR ACETABULAR SYSTEM, CUP, HEMISPHERICAL, MULTI HOLE, 64
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
5128346313
MDR Report Key22303250
Report Number1644408-2025-00851
Device Sequence Number15442098
Product Code LPH
UDI-Device Identifier00888912144599
UDI-Public00888912144599
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K190057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number940-03-64J
Device Lot Number821Z1004A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2025
Initial Report FDA Received Date06/20/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
940-00-015 LOT: 144Z1108; 940-00-020 LOT: 145Z1151; 940-00-025 LOT: 146Z1296; 940-00-030 LOT: 147Z1226; 940-00-035 LOT: 148Z1237; 940-00-045 LOT: 150Z1053; 951-01-54J LOT: 2221A1008; 952-28-54J LOT: 2229A1017
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
-
-