• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*EXTRA TITANIUM CLIPS; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*EXTRA TITANIUM CLIPS; CLIP, IMPLANTABLE Back to Search Results
Catalog Number LT100
Medical Device Problem Codes Use of Device Problem (1670); Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent 6/23/2025, d4 batch # unk.An analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.However, if the product is received at a later date, the investigation will be updated as applicable.Photo images were received.Any additional information obtained from the photo evaluation will be included as part of the product investigation.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a cardiovascular procedure the cardiac surgery claims that during the surgical procedure the clip does not work correctly, it does not grip as it should, the legs of the clip cross and during the procedure the clip comes loose.The clipper is not johnson.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Date sent 6/30/2025.Photo analysis: this is an analysis of three photos submitted for evaluation.During the visual analysis, the following was observed: first photo shows a package from label side with product code lt100, lot # 267d91, expiration date: 2029-08-31.Since the picture don¿t have sufficient evidences to added a potential cause.Second photo shows a device from bhiosupply.Third photo shows a device from jaws area with s/n (b)(6).Based on the photos the event described cannot be confirmed.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.A manufacturing record evaluation was performed for the finished device lot number 267d91, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
3035526892
MDR Report Key22306607
Report Number3005075853-2025-04679
Device Sequence Number20853443
Product Code FZP
UDI-Device Identifier10705036012788
UDI-Public(01)10705036012788(17)290831(10)267D91
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K830503
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberLT100
Device Lot Number267D91
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/19/2025
Supplement Date Received by Manufacturer06/27/2025
Initial Report FDA Received Date06/23/2025
Supplement Report FDA Received Date06/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-