| Catalog Number |
LT100 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 6/23/2025, d4 batch # unk.An analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.However, if the product is received at a later date, the investigation will be updated as applicable.Photo images were received.Any additional information obtained from the photo evaluation will be included as part of the product investigation.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a cardiovascular procedure the cardiac surgery claims that during the surgical procedure the clip does not work correctly, it does not grip as it should, the legs of the clip cross and during the procedure the clip comes loose.The clipper is not johnson.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 6/30/2025.Photo analysis: this is an analysis of three photos submitted for evaluation.During the visual analysis, the following was observed: first photo shows a package from label side with product code lt100, lot # 267d91, expiration date: 2029-08-31.Since the picture don¿t have sufficient evidences to added a potential cause.Second photo shows a device from bhiosupply.Third photo shows a device from jaws area with s/n (b)(6).Based on the photos the event described cannot be confirmed.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.A manufacturing record evaluation was performed for the finished device lot number 267d91, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|