|
Correction: e3 the unit was initially evaluated in the field by a for medical field service technician.The reported complaint description is confirmed.The unit was opened and the power supply contained 2 blown capacitors.The power supply module was replaced and the unit functioned as intended.The root cause for the "burnt/smoke" was determined to be caused by two blown capacitors in the power supply module, which was replaced.This is the first reported error of this unit for burnt.This is the first time the power supply module has been replaced.The unit was tested with the new power supply module per svc-002-s10 functional test and svc-027 electrical safety test and met all acceptance criteria.A review of the device history records (service order history) was performed for the reported serial number 07040724 for any deviations related to the reported failure mode of the complaint.The review confirmed that the unit met all material, assembly, and performance specification prior to distribution; i.E.No ncr associated with reported failure mode.Labeling review: the user manual (nanoknife user manual, {16755922-21), which is supplied to the user with this unit states: section 5: system operation 5.1 procedure overview an overview of a typical nanoknife ablation procedure is listed below.Refer to subsequent sections of this user manual for detail usage instruction on operation of the nanoknife generator.5.1.1 procedure setup (before patient enters procedure room): 1.Plug in the nanoknife generator and cardiac gating device into a grounded power outlet within the procedure room.2.Power on the nanoknife generator.The nanoknife generator will initiate and complete a power-on self-test (post).3.Attach the double pedal footswitch to the nanoknife generator.5.1.2 patient preparation 4.Prepare patient for general anesthesia.5.Position patient into appropriate position for anticipated nanoknife single electrode probe insertion (e.G., supine, prone, lateral, lithotomy)."maintenance should be carried out only by trained personnel.The generator must undergo periodic preventative maintenance as specified in the maintenance and service (section 9).Avoid moving the device when powered on.Avoid jarring the equipment during transport.System location: the generator must be installed and operated in an environment that conforms to the operating conditions specified in section 10.4.The generator must be installed on rigid surfaces suitable to withstand its weight.Maintenance calibration required every 12 months by an authorized service agent." a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|